If you are a college or university looking to setup a complete lab. Please contact us with your details for custom quotation. sales@aticoexport.com

Automatic Dropping Point Apparatus of Lubricating Grease Over Wide Temperature Range

We are the leading manufacturer and supplier of Automatic Dropping Point Apparatus of Lubricating Grease Over Wide Temperature Range. Now accepting bulk orders for government tenders in every major country around the globe.

Use:

  • The dropping point of a grease is the temperature at which it passes from a semi-solid to a liquid state. The dropping point test determines the cohesiveness of the oil and thickener of a grease, or when the thickener breaks down. Dropping point tests are used for quality control to verify that the soap structure has been formed correctly and will withstand high temperatures.
  • Does not apply for PFPE/PTFE thickened greases

Compliance with following International Standards:

  • Dropping Point is also known as ASTM Method Automatic Dropping Point Apparatus.
  • Automatic Dropping Point Apparatus Standard Test Method for Dropping Point of Lubricating Grease Over Wide Temperature Range.
  • The equipment is used for higher temperature grease testing than the ASTM D566 method. The different heating procedure is used as compared with ASTM D566.

Feature:

  • 7-inch LCD touch screen HMI with beautiful outlook design. Full English operating system is stalled. Instruction interfaces are clearly displayed to provide guide for test completion.
  • High-performance microprocessor as core of MCU, fully automatic control throughout the whole test procedures.
  • Highly integrated bench top instrument with multiple functions.
  • Specially designed photoelectric sensor for dropping point detection with high sensitivity and fast response.
  • It features double sample holes, single test or dual tests can be performed.
  • Fast temperature rising rate with ±1℃ metal bath temperature accuracy.
  • It features fault self-diagnosis system and reminder functions.
  • Automatic test results and experiments number saving functions with auto test results printing.
  • User-friendly and easy-to-operate, average test time is less than 2 minutes.

Specifications:

  • Operating voltage: AC220V, 50Hz
  • Operating power: 1200W
  • Metal bath type: Aluminum
  • Sample channels: 2 holes
  • Temperature sensor: Class A PT100*2
  • Test method: auto photoelectric sensor
  • Temperature control range: R.T. ~ 400℃
  • Temperature control point: 121℃, 232℃, 288℃, 316℃, 343℃
  • Temperature control accuracy: ±1℃
  • Test result capacity: automatic saving 256 test results
  • Printing method: auto thermal printer

 

 

 